3. SPECIFICATION
Function
Range
Resolution
Accuracy
AC Voltage: (Auto-ranging except 200mV)
Input Impedance: 10MΩ
Frequency Range:50 to 60Hz
Maximum Input: 600V dc or 600V ac rms (CATIII)
Maximum Input: 1000V dc or 1000V ac rms (CATII)
200.0mV
0.1mV
±1.5% reading ± 30 digits
2.000V
1mV
±1.2% reading ± 3 digits
20.00V
10mV
±1.5% reading ± 3 digits
200.0V
100mV
±1.5% reading ± 3 digits
600V
1V
±2.0% reading ± 4 digits
1000V
1V
±2.0% reading ± 8 digits
Function
Range
Resolution
Accuracy
DC Voltage: (Auto-ranging)
Input Impedance: 10MΩ
Maximum Input: 600V dc or 600V ac rms (CATIII)
Maximum Input: 1000V dc or 1000V ac rms (CATII)
200.0mV
0.1mV
±0.5% reading ± 2 digits
2.000V
1mV
±1.2% reading ± 2 digits
20.00V
10mV
200.0V
100mV
600V
1V
±1.5% reading ± 2 digits
1000V
1V
±1.5% reading ± 4 digits
Function
Range
Resolution
Accuracy
DC Current: (Auto-ranging for uA / mA)
Overload Protection: 0.2A / 250V and 10A / 250V
Fuse.
Maximum Input: 200mA dc or 200mA ac rms on uA /
mA ranges, 10A dc or ac rms on 10A range.
200.0µA
0.1µA
±1.0% reading ± 3 digits
2000µA
1µA
±1.5% reading ± 3 digits
20.00mA
10µA
200.0mA
100µA
10A
10mA
±2.5% reading ± 5 digits
Function
Range
Resolution
Accuracy
AC Current (Auto-ranging for uA / mA)
Overload Protection: 0.2A / 250V and 10A / 250V
Fuse.
Frequency Range: 50 to 60Hz
Maximum Input: 200mA dc or 200mA ac rms on uA /
mA ranges, 10A dc or ac rms on 10A range.
200.0µA
0.1µA
±1.5% reading ± 5 digits
2000µA
1µA
±1.8% reading ± 5 digits
20.00mA
10µA
200.0mA
100µA
10A
10mA
±3.0% reading ± 7 digits
Function
Range
Resolution
Accuracy
Resistance (Auto-ranging)
Input Protection: 500V dc or 500V ac rms.
200.0Ω
0.1Ω
±1.2% reading ± 4 digits
2.000KΩ
1Ω
±1.0% reading ± 2 digits
20.00KΩ
10Ω
±1.2% reading ± 2 digits
200.0KΩ
100Ω
2000MΩ
1KΩ
20.00MΩ
10KΩ
±2.0% reading ± 3 digits
Function
Test Current
Audible Continuity Test
Audible threshold: Less than100Ω
Overload protection: 500V dc or ac rms.
0.3mA typical
NOTE!Accuracy specifications consist of two elements:
1.
(% reading) – This is the accuracy of the measurement circuit.
2.
(± digits) – This is the accuracy of the analog to digital converter.
NOTE! Accuracy is stated at 65°F to 83°F (18°C to 28°C) and less than 70% RH (Relative Humidity).