beautypg.com

Sealey TA300 User Manual

Page 2

background image

Item

Description (Fig.1)

1.

“HOLD” button.

After “Hold” button is pressed, the present reading will be held in the display.

2.

“LIGHT” button.

To enable spotlight function, press “Light” button.
To disable, press the “Light” button again.

3.

“SELECT” button.

When meter is in the Vdc range press the “Select” button to change to
the Vac range, press “Select” again and the meter will return to the Vdc range..
When the meter is in the resistance range press the “Select” button to change to

to diode range. Press “Select” again, for continuity range. Press “Select”

again for capacitance range.
When the meter is in frequency range, press the “Select” button for
duty cycle.

4.

“POWER” button.

The “Power” button is used for switching the meter on.

Press “Power” again to turn the meter off.

5.

“FUNCTION” switch.

Use “Function” switch to select between Vac, Vdc, resistance,
capacitance, frequency, duty cycle, continuity test, audible continuity
and diode test.

6.

Test probe. (+Test Probe)

7.

Test lead.

(-Test Lead)

8.

Crocodile clip.

9.

Cigarette lighter adaptor.

1

2

3

5

6

4

7

8

9

DC VOLTAGE

Range Resolution

Accuracy

400mV

0.1mV

4V

1mV

± (0.8%+4)

40V

10mV

400V

100mV

600V

1V

± (1%+4)

RESISTANCE

Range Resolution

Accuracy

400

0.1

± (1.2%+2)

4k

1

40k

10

± (1%+2)

400k

100

4M

1k

± (1.2%+2)

40M

1

0k

± (1.5%+2)

CAPACITANCE

Range Resolution

Accuracy

4nF

1pF

± (25%+5)

40nF

10pF

400nF

100pF

4µF

1nF

± (10%+5)

40µF

1nF

100µF

100nF

DUTY CYCLE

Range Resolution

Accuracy

0.1% - 99%

0.01%

± (2.5%+4)

DIODE

Range Resolution

Open-circuit Voltage:

1mV

About 1.5V.
Forward voltage:
About 0.5 - 0.8V

FREQUENCY

Range

Accuracy

9.999Hz
99.99Hz
999.9Hz

± (1.5%+4)

9.999kHz
99.99kHz
999.9kHz
10MHz

AC VOLTAGE

Range Resolution

Accuracy

4V

1mV

40V

10mV

± (0.8%+6)

400V

100mV

600V

1V

± (1.2%+6)

CONTINUITY

If < 50

, the buzzer will sound

Accuracy is specified for a period of one year after calibration and at 23 ± 5°C with relative humidity up to 75%.
Accuracy specifications take the form of:
± (% of Reading + number of least significant digits).

Input impedance: 10

Overload protection: 250Vdc or rms ac

Input voltage range: 0.5 - 50 Vp-p

Note: When the resistance is between 50

and

120

Ω,

the buzzer may or may not sound. When

the resistance is more than 120

, the buzzer

will not sound.

Input voltage range: 0.5 - 50 Vp-p

Overload protection: 250Vdc or rms ac

Input impedance: 10M

Frequency response: 40 - 400\Hz

3.

SPECIFICATION

TA300 - 1 - 310807

Fig.1