Specification – Sealey TA201 User Manual
Page 2
Range Resolution Accuracy
400 .0mV 0 .1mV
± 0 .5% of reading ± 2 digits
4 .000V 1mV
40 .00V
10mV
±1 .5% of reading ± 2 digits
400 .0V 100mV
1000V
1V
± 1 .8% of reading ± 2 digits
Range Resolution Accuracy
400 .0mV
0 .1mV
± 1 .5% of reading ± 30 dig
4 .000V
1mV
± 1 .0% of reading ± 3 digits
40 .00V
10mV
± 1 .5% of reading ± 3 digits
400 .0V 100mV
600V
1V
± 2 .0% of reading ± 2 digits
Range Resolution Accuracy
400 .0uA
0 .1uA
± 1 .0% of reading ± 3 digits
4000uA 1uA
40 .00mA
10uA
± 1 .5% of reading ± 3 digits
400 .0mA 100uA
4A
1mA
± 2 .5% of reading ± 5 digits
10A 10mA
Range Resolution Accuracy
400 .0Ω
0 .1Ω
± 1 .2% of reading ± 4 digits
4 .000kΩ
1Ω
± 1 .0% of reading ± 2 digits
40 .00kΩ
10Ω
± 1 .2% of reading ± 2 digits
400 .0kΩ
100Ω
4 .000MΩ
1kΩ
40 .00MΩ
10kΩ
± 2 .0% of reading ± 3 digits
Range Resolution Accuracy
400 .0uA
0 .1uA
± 1 .5% of reading ± 5 digits
4000uA 1uA
40 .00mA
10uA
± 1 .8% of reading ± 5 digits
400 .0mA 100uA
4A
1mA
± 3 .0% of reading ± 7 digits
10A 10mA
DC Voltage (Auto Ranging)
AC Voltage (Auto Ranging except 400mV)
Input Impedance: 10MΩ.
Maximum Input 600Vac rms or 600Vdc .
DC Current (Auto Ranging for uA and mA)
Input Impedance: 10MΩ.
Frequency Range: 50 to 400Hz .
Maximum Input: 600Vac rms or 600Vdc .
Overload Protection: 0 .5A/250V and 20A/250V
Fuse:
Maximum Input:400mAac rms or 400mAdc on uA/mA ranges
20A ac or dc rms on 10A range .
AC Current (Auto Ranging for uA and mA)
Overload Protection: 0 .5A/250V and 20A/250V
Fuse:
Frequency Range:50 to 400Hz
Maximum Input: 40mAac or 400mAdc rms on uA/mA ranges,
20Aac rms or dc on 20A range .
Resistance (Auto Ranging)
Range Resolution Accuracy
40 .00nF
10pF
± 5 .0% of reading ± 7 digits
400 .0nF 0 .1nF
4 .000uF
1nF
± 3 .0% of reading ± 5 digits
40 .00uF 10nF
100 .0uF
0 .1uF
± 5 .0% of reading ± 5 digits
Capacitance (Auto Ranging)
Input Protection: 250Vac rms or 250Vdc .
Input Protection: 250Vac rms or 250Vdc .
Range Resolution Accuracy
5Hz 0 .001Hz
50Hz
0 .01Hz
± 1 .5% of reading ± 5 digits
500Hz 0 .1Hz
5kHz 1Hz
50kHz
10Hz
± 1 .2% of reading ± 3 digits
500kHz 100Hz
10MHz
1kHz
± 1 .5% of reading ± 4 digits
Frequency (Auto Ranging)
Sensitivity: <0 .5V rms while <1MHz
Sensitivity: >3V rms while >1MHz
Overload Protection: 250Vac rms or dc .
Range Resolution Accuracy
0 .1%~99 .9%
0 .1%
± 1 .2% of reading ± 2 digits
Duty Cycle (Auto Ranging)
Pulse Width: >100us, <100ms
Frequency Width: 5Hz - 150kHz
Overload Protection: 250Vac rms or dc .
Range Resolution Accuracy
-20°C~+760°C 1°C
± 3 .0% of reading ± 3 digits
-4°F~+1400°F 1°F
(Meter only, probe accuracy
not included) .
Range
Resolution Accuracy
RPM 4 600~4000RPM
1RPM
600~1200RPM
10RPM
(X 10 RPM)
±2% of reading ± 4
RPM 2 300~4000RPM
1RPM
digits
/DIS 300~600RPM 10RPM
(x 10 RPM)
Temperature
Sensor: Type K Thermocouple
Range Resolution Accuracy
0 .3mA
1mV
± 10% of reading ±5 digits
typical
Diode test
Open Circuit Voltage: 1 .5Vdc typical
Overload Protection: 250Vac rms or dc .
Cylinder Range Resolution Accuracy
2CYL 0~180 .0°
3CYL 0~120 .0°
4CYL
0~90 .0° 0 .1°
± 2 .0% of reading ± 4 digits
5CYL 0~72 .0°
6CYL 0~60 .0°
8CYL 0~45 .0°
10CYL 0~36 .0°
Dwell Angle
Overload Protection: 250Vac rms or dc .
Effect Reading: >600RPM
Overload protection 250V ac rms or dc .
Overvoltage Category:
CAT111 600V .
Display:
4000 counts LCD display with function indication .
Polarity:
Automatic, (-) negative polarity indication .
Overrange:
“OL” mark indication .
Low Battery Indication:
The “BAT” is displayed when the battery voltage drops below the operating level .
Measurement Rate:
2 times per second nominal
Auto Power Off:
Meter automatically shuts down after approx 30 minutes of inactivity .
Operating environment:
0°C to 50°C (32° to 122°F) at < 70%relative humidity .
Storage Temperature:
-20° to 60°C (-4°F to 140°F) at <80% relative humidity .
For Inside Use:
Max Working Height:
2000mtr .
Power:
One 9V battery PP3 .
Dimensions:
146 (H) x 66 .2 (W) x41 .5 (D) mm
Approx Weight:
330g
Accuracy is given at 18°C to 28°C (65°F to 83°F) less than 70% relative humidity.
RPM (Tacho)
Audible Continuity
Audible threshold:
Less than 150Ω
Test Current:
<0 .3mA
Overload Protection: 250Vac rms or dc .
Original Language Version
3. SPECIFICATION
© Jack Sealey Limited
TA201 Issue No: 3(L) - 17/06/14