Initial Appraisal
A number of accessories accompany the Hydrosteel 6000 Analysis Kit. When you first receive your Analysis
Kit, we suggest that you check the items received against the Packing List. If any item is damaged or
missing, notify your supplier, stating the serial number stamped on the rear of the handheld analyser,
together with your order number and the supplier’s invoice number. Familiarise yourself with the components
of the Hydrosteel 6000 Analysis Kit by going to the ‘Contents of the Hydrosteel 6000 Complete Kit’.
System Assembly and Start-up
Normalising the analyser
1
Remove the analyser from the field case.
2
Remove the leather outer case.
3
Using the special key provided, open the battery compartment (at the rear of the analyser) and insert
the four batteries complete with their holder (supplied)
– see the diagram below.
Close and lock the battery compartment.
4
Refit the leather outer case. Press the on/off key (I/O), which is located on the analyser front panel.
5
A countdown shows the time remaining, and a red LED flashes. The detector is then normalised,
and a hydrogen flux measurement of (default units of pl/cm
2
/s) is displayed.
6
Your Hydrosteel 6000 analyser is now operational.
7
When not required for immediate use, switch off the analyser by pressing the on/off key (I/O). The
analyser takes a few seconds to shut down, during which time you can abort the switch-off by again
pressing the I/O key.
WARNING
The analyser batteries must be alkaline-type AA-
size. Using any other type of battery will
invalidate the Intrinsic Safety approval.
Use the special key to
open/close the battery
compartment