beautypg.com

Ion Science Hydrosteel 6000 User Manual

Page 15

background image

Hydrosteel 6000 MANUAL

Ion Science Ltd

Page 15 of 39

Unrivalled Detection. www.ionscience.com

Using Hydrosteel 6000

Probe indication
To the left of the battery icon the display one of the following, denoting the probe in use. Ensure at all times
that the probe in use and probe indicator correspond. If not, reconfigure the analyser to indicate the correct
probe from the analyser start up menu, as indicated in Appendix B. The probe indicator should correspond
to the probe in use as follows:

LT


LT-R (low temperature roaming)

HT


HT-R (high temperature roaming)

(blank)


ambient hydrogen measurement
(probe independent)



Using an External Power Supply
To extend the continuous operation time of the analyser, particularly for long term monitoring applications, an
external power supply can be used.

For satisfactory operation, the analyser requires a 6 V dc regulated power supply, capable of delivering 200
mA minimum.

Note: It is important to ensure that the centre pin of the connector is positive with respect to the shroud.

Failure to provide this connection may cause damage to the analyser circuitry.





























WARNING

If the Hydrosteel analyser is to be used in an intrinsically safe zone, any

external power supply must have appropriate safety certification.