Routine maintenance

1) Open the regulator. By turning the valve knob as indicated.


















2) Connect the Hydrosteel 6000 analyser to the 2 ft LT-R sample tube then connect the end of the

sample tube to the flood leak as shown .











3) Leave the instrument connected for 1 to 2 minutes for the reading to stabilise.
4) For the LT probe the readings should be between 280 pl/cm

2

/s and 380 pL/cm

2

/s. For the HT probe

the between 2200 pL/cm

2

/s and 3000 pL/cm

2

/s.