Routine maintenance
1) Open the regulator. By turning the valve knob as indicated.
2) Connect the Hydrosteel 6000 analyser to the 2 ft LT-R sample tube then connect the end of the
sample tube to the flood leak as shown .
3) Leave the instrument connected for 1 to 2 minutes for the reading to stabilise.
4) For the LT probe the readings should be between 280 pl/cm
2
/s and 380 pL/cm
2
/s. For the HT probe
the between 2200 pL/cm
2
/s and 3000 pL/cm
2
/s.