beautypg.com

Position encoders and reference marks – HEIDENHAIN NPD 281 v.3 User Manual

Page 6

background image

6

Scale in

Distance-coded

linear encoder

reference marks

Position Encoders and Reference Marks

The ND 281 and NDP 281 display units are primarily intended
for use with photoelectrical linear encoders with sinusoidal
11-µA

PP

or 1-V

PP

signals: primarily for HEIDENHAIN

MT length gauges with 11 µA

PP

.

These length gauges have one reference mark. The scales of
other photoelectric linear encoders (see “Linear Encoders”)
can contain one reference mark or several

distance-coded

reference marks.

If there is an interruption of power, the relationship between
the position of the length gauge and the displayed position
value is lost. The reference marks on the position encoders
and the REF reference mark evaluation feature enable the
display unit to quickly re-establish this relationship again
when the power is restored.

When a reference mark is crossed over, a signal is generated
which identifies that position as a reference point. At the
same time, the display unit restores the relationship between
length gauge position and display values which you last
defined by setting the datum.

If the linear encoders have distance-coded reference marks,
you only need to traverse a maximum of 20 mm to restore
the datum.

Reference mark

Reference marks on linear encoders

Position Encoders and Reference Marks