Switching Inputs EXT (X41)
Weg
5
t
t
t
v
t
v
Pin 15
(A1)
Time curve of a signal at pin 15 for trigger limit (A1) = 5 mm , t
d
≤
22 ms
Zero crossover
The display value “zero” activates the corresponding output
(pin 14). Minimum signal duration is 180 ms.
Path
Output
Measured value < Lower sorting limit
Measured value > Upper
sorting limit
Path
Lower limit Upper limit
Switching point
Switching signals
As soon as the trigger points defined in parameters are
reached, the corresponding outputs (pins 15,16) are activated.
You can set up to two trigger points. The switching point
“zero” has a separate output (see “Zero crossover”).
Signals for sorting and tolerance checking
If the sorting limits defined in parameters are exceeded, the
corresponding outputs (pins 17, 18) are activated.
Signals
Operating parameters
Pin
Switching signals
P62, switching limit 1
15
P63, switching limit 2
16
Sorting signals
P18, lower sorting limit
17
P19, upper sorting limit
18
t
d
t
d
Path