List of Operating Parameters

Parameter

Settings / Function

Linear error compensation
– 99 999,9 < P41 < + 99 999,9 [µm/m]

Factory setting:

0

Example: Determine input value for P41
Displayed length ............................... L

d

= 620.000 mm

Actual length (as determined,
for example, with the VM 101
comparator system
from HEIDENHAIN) ........................... L

a

= 619.877 mm

Difference .........................................

L = L

a

– L

d

= – 124 µm

Compensation factor k (= P41):
k =

L / L

d

= – 123 µm / 0.62 m ....... k = – 198,4 [µm/m]

Reference marks
One reference mark

Distance-coded with 500 • SP
(SP: signal period)

Distance-coded with 1000 • SP
(e.g. for HEIDENHAIN LS ...C)

Distance-coded with 2000 • SP

Distance-coded with 5000 • SP

Reference mark evaluation
Evaluate reference marks

Do not evaluate
reference marks

Parameter

Settings / Function

Encoder monitoring
Monitoring not active

Contamination

Frequency

Contamination and frequency

Baud rate
110 / 150 / 300 / 600
1200 / 2400 / 4800 / 9600 baud

Additional blank lines for
data output
0

P51

99

Factory setting: 1

Trigger limit 1

Trigger limit 2

Value for datum point
Enter numerical value for
datum setting over switching input
or with ENT key