RS-232-C/V.24 Interface (X31)

baud rate

Measured value output after signal through the “Contact”
or “Pulse” inputs

To start measured value output through the EXT interface
(X41) you can either:

➤ Close the “Contact” input (pin 23 on X41) against 0 V,

for example with a simple switch (make contact);
or

➤ Close the “Pulse” input (pin 22 on X41) against 0 V,

for example by triggering the input with a TTL logic
device (such as SN74LSxx).

Characteristic times for measured value output

Process

Time

Minimum duration of “Contact” signal

t

e

7 ms

Minimum duration of “Pulse” signal

t

e

1.5 µs

Storage delay after “Contact”

t

1

5 ms

Storage delay after “Pulse”

t

1

1 µs

Measured value output after

t

2

57 ms

Regeneration time

t

3

0

The time for measured value output (t

2

) is the

longest during a DIFF series of measurements.

Duration of measured value transfer

t

D

=

[s]

Triggering the “Contact” and “Pulse” inputs at D-sub connection
EXT (X41)

Signal transit times for measured value output after “Pulse” or “Contact”

Pin 1(0V)

Pin 23

Pin 22

Pin 1(0V)

EXT(X41)

EXT(X41)

t

1

t

2

t

e

t

e

t

D

t

3

187 + (11 • number of blank lines)