beautypg.com

Power up -1, Start up oxymitter 4000 calibration -1, Hazardous area oxymitter 4000 – Emerson Process Management OXYMITTER 4000 User Manual

Page 79: O2: 2.59% lk normal

background image

Instruction Manual

IB-106-340C Rev. 4.1

July 2004

Rosemount Analytical Inc. A Division of Emerson Process Management

Startup and Operation with LOI 6-1

Hazardous Area Oxymitter 4000

SECTION 6

STARTUP AND OPERATION

OF HAZARDOUS AREA OXYMITTER 4000 WITH LOI

6-1

POWER UP

a. Startup

Display

When power is applied to the probe, the cell

heater turns on. It takes approximately one
half hour for the cell to heat to operating
temperature. This condition is indicated by a
“warm up” display on the LOI (Figure 6-1).
This message will continue to display until
the cell is up to operating temperature.

b. Operating

Display

The normal operating display is the % O

2

concentration. The “normal” display is
shown in Figure 6-2.

c. Error

If there is an error condition at startup, an

alarm message will be displayed. Refer to
Section 8, TROUBLESHOOTING, to deter-
mine the cause of the error. Clear the error,
cycle power, and the % O

2

display should

return.

d. LOI

The Local Operator Interface can be used to

change the software and alarm settings, to
adjust the high and low gas settings, and to
initiate the calibration sequence. Refer to
the LOI menu (Figure 6-4).

e. Reference

Air

Ensure the reference air, if used, is set to

56.6 l/hr (2 scfh).

6-2

START UP OXYMITTER 4000
CALIBRATION

Refer to Section 9, MAINTENANCE AND
SERVICE, for calibration instructions.

37260035

O2:

0.00%

LK

warm up

367dgC

Figure 6-1. Startup Display

37260036

O2:

2.59%

LK

normal

Figure 6-2. Normal Display

6