beautypg.com

Hazardous area oxymitter 4000 – Emerson Process Management OXYMITTER 4000 User Manual

Page 124

background image

Instruction Manual

IB-106-340C Rev. 4.1
July 2004

9-6 Maintenance and Service

Rosemount Analytical Inc. A Division of Emerson Process Management

Hazardous Area Oxymitter 4000

4. Touch the Enter arrow to start the Gas

2 flow. The timer is activated and the
LOI displays:
Flow Gas 2xxxxs
Read Gas 2xxxxs
Done Gas 2
Stop Gas
Hit E when ready

5. Remove the second calibration gas

and cap off the calibration gas port.
Then, touch the Enter arrow to indicate
completion. The timer is activated and
the LOI displays:
Purgexxxxs

The default purge time is three min-
utes. When the gas purge timer times
out, the Oxymitter 4000 begins to read
the process O

2

.

b. Abort

Calibration

Exits the calibration. After calibration gases

are removed, and the purge times out, the
instrument goes back to normal operational
mode.

c. Cal Constants – Results of the

Calibration

1. Current calibration – If the calibration

passed these values will be updated.
Log these values onto the calibration
log sheet supplied. If the process has
high levels of particulate, the response
back to the process after cal gas is
also removed.

2. Previous Calibration – Values from the

prior good calibration.

3. Failed Calibration – Bad calibration

values are not loaded into the elec-
tronics.

d. Calibration

Status

1. Calibration Step – The current step in

an active calibration procedure.

2. Calibration Time – Time until the next

scheduled calibration.

3. Next O2 Cal – Time until the next O

2

calibration, if different than the next
scheduled calibration.