Operators and Reference Manual

MC12

MC12 OM RM 1.03_EN

Version: 1.03

Date: 2010-11-17

Page 35 of 46


Enter the required format followed by and the system will apply the chosen flow rate
format for read outs in Function 12.

6.7.47. Function 72 - Volume format

Changes the unit for F1 and F11
Volume format for function 1 and 11.
The function makes it possible to change volume format between:
1: volume (ml) and 2: weight (g).
This is effective in function 1 and 11 as well as in print out.
When the value 2: weight (g) is chosen, the F9 is automatically set to 1.0000 g/ml and can not be
changed.

6.7.48. Function 73 - Display MC12 version

This function displays the type, version and date of the MC12 software.
When used type and version will remain visible until is entered.

6.7.49. Function 80 - Reset memory

This function resets the memory with the exception of the part used for saving programs.
MC12 will enter stand-by mode, and when switched on again the built-in parameters will be valid.

IF MC12 DOES NOT WORK OR DOES NOT OPERATE AS EXPECTED,
ACTIVATE FUNCTION 80

A calibration is necessary after using Function 80.

6.7.50. Function 86 - Complete memory reset

This function works in the same manner as Function 80 but includes also the part of the memory
containing programs saved via Function 31.

This function should only be used for extreme cases.
Function 86 is not shown in the status lines of the display.

A calibration is necessary after using function 86.