beautypg.com

Operators and reference manual mc12, Programming – Watson-Marlow MC12 User Manual

Page 10

background image

Operators and Reference Manual

MC12

MC12 OM RM 1.03_EN

Version: 1.03

Date: 2010-11-17

Page 10 of 46

6. PROGRAMMING

6.1 Starting the MC12


In the following, the sign <> will mean that the indicated key must be activated. For instance,
means that ENTER must be pressed.

When the MC12 is connected, the main switch on the left side of the control box can be turned
on. The display will show the following:

STANDBY





Figure 5 – Starting MC12 -I

Press the key, and the display will show the following:

* MC12 V1.XX (C) Flexicon 90-XX *
PRESS GO TO CONTINUE.

DRIVE No.: 1

Figure 6 – Starting MC12 -II


Verify that all connected drives are displayed with their respective numbers.
Then press once more, and the display will show the following:

ENTER FUNCTION NO. :
F 1: 100.00 ml.
F 3: 200 rpm
F 5: 0 REV.

I1

F 2: 8.0 TUBE
F 4: 10 ACC
F 6: 1 FILLS

Figure 7 – Starting MC12 -III


The values shown in the status lines will be the above or the latest values used.

Please note – that in MC12 special versions, pressing the GO key during start up is put out of
commission. Start-up runs without pressing GO.

6.2 Password Protection


MC12 is password protected. To be able to perform the following programming procedures it is
necessary to log on as a supervisor (see function 59).
Always remember to change the password level back to user level after programming.

Remember that merely switching MC12 off/on will NOT change any selected password level!