6.3 Parameters


In the following a parameter will be a single value like for example volume, velocity, number of
fillings etc.

6.4 Programs


In the following a program will be a complete set of parameters which together form a program
according to which MC12 and the connected pump will operate.

6.5 General information on the programming of MC12


MC12 is equipped with a battery in the memory and will therefore always remember the
programmed parameters, even if the main switch is turned off.

The programming is made via functions, i.e. every operating parameter has its own function
number.

The programming is made by entering the function number followed by "ENT".
This will make the required function appear in the prompt line of the display and show the current
value or information of the function.

This value will automatically be overwritten when entering the new value. After being entered, the
new value will be displayed in the prompt line. The new value is entered into the computer by
pressing "ENT".

The new value will be displayed in the status lines at once.

Example:
If a volume of 8.5 ml is required, the following should be entered:
<1>++<8>+<.>+<5>+

MC12 knows automatically which filler is connected. Consequently, values which are not valid for
this pump will not be accepted.

If a faulty value is entered, it will not be entered, and a "beep" sound will be heard.