Print audit routines

If a printer is connected, the completion of a dose run will automatically call the print

routine. The first request will be to enter the operator ID.

Up to 16 characters can be entered. Digits and the decimal point are entered directly from

the keypad. Alpha characters are entered by pressing Up or Down which call A to Z and Z

to A respectively in circular rotation.

An alpha character is embedded by pressing Step. A numeric character is entered by

pressing Step, any other numeric character, the decimal point or Up or Down.

On pressing Enter, the pump will request the input of a batch number.

Again, up to 16 characters can be entered as for the operator ID. When Enter is again

pressed, the following information is printed out: date, time, dose size, specific gravity,

dose interval, number of doses, initial ml/rev, recalibration data, operator ID, batch

number and number of doses delivered.

Following the print-out a repeat dose option will be given.

Single dose command

Single doses can be dispensed on demand, with a count of the number of doses being

kept.

Set the interval time to 0 SECONDS and the number of doses to 1.

To start single dosing, press Start or use an external start dose switch. The display will

indicate the total number of completed doses up to a maximum of 99,999, after which the

counter will restart at 0: dose 100,001 would be shown as 1.

Calibration procedure

Calibration of the 624Di is based on informing the pump under Cal of the pumphead and

tubing which are to be used. Alternatively a calibration dose can be used. The calibration

dose will run for a maximum of 4 minutes, but can be stopped at any time up to 4

minutes. The longer the calibration dose, the more accurate the calibration. Entering into

the pump the physical volume or mass to complete the procedure will allow the pump to

take into account ambient conditions and also the viscosity of the fluid.

Manual operation

Switch power on.

Change the set speed by pressing the Up or Down key. The minimum speed of the

624Di is 4 rpm.

Change direction by pressing the CW/CCW key. Indication of direction of rotation is

provided via the LCD display.

Select the maximum speed: press the Up key and the Max key together. Select the

minimum speed: press the Down key and the Max key together.

Press Start to start the pump. Press Stop to stop the pump.

Auto operation

The pump will accept external control signals through the 25-pin cage clamp connector on

the back panel. Remove the cover plate, ensuring that the gasket is not damaged. Feed

the control wires through the cable glands and connect via the sprung cage clamps.

Analogue

This function enables the pump speed to be controlled via an external analogue process

signal. Pressing Enter at analogue will call a confirmation of the analogue control signal

settings. These can be reset under Setup (see section covering pump setup).

RS232

This facility gives full pump functionality under RS232 closed loop control via the 4-pin

cage clamp. Up to 16 pumps can be linked while retaining individual pump control by using

lead PR0036. A network kit is available from Watson-Marlow which includes Pumpnet 2, a

DOS-compatible control program, and leads.