9.2.3 Open the meter case by wiggling the back to separate it from
the front. Set the back of the case to one side.

9.2.4 Locate the defective fuse and replace with the exact or
equivalent type. See meter specifications.

9.2.5 Reassemble case of meter. Reinstall screws.

9.2.6 Verify that the meter operates properly before using it to
make measurements.