2.34

Do not use this meter to measure current in circuits whose

open circuit voltage exceeds 250V AC/DC. The meter’s fuses
are rated at 250V max. Failure to observe this precaution
may result in damage to the meter or injury to the user.

2.35

When you use the meter to check a high-voltage circuit, do

not try to connect both test leads at once. Instead, clip one

probe to the neutral or ground lead of the circuit (usually

a bare, green, or white wire in AC wiring circuits) using the

insulated slip-on Alligator Clips. Then probe for voltages

with the other test lead. This helps prevent you from

accidentally touching a hot wire, since you need to

concentrate on only one test lead. Failure to observe

this precaution may result in damage to the meter or

injury to the user.

2.36

If there is any doubt about the condition of the meter

(i.e. safe vs unsafe), remove the meter from service and

secure it in a location that will prevent its unintentional use.

Failure to observe this precaution may result in damage to

the meter or injury to the user.