beautypg.com

5 outputs – Tecfluid CU Series User Manual

Page 29

background image

29

6

When the read flow rate is stable, press the key OK and the instrument will save the
value.
The following step is to swap the cables of the transducers: the cable TR1 for the
transducer TR1 has to be connected to the transducer TR2 and vice versa.

Do not move the transducers from its position. Only unscrew the two connectors and
swap the cables as described.

Once a key is pressed, the following screen is shown:

After pressing OK, the cables must be placed in its original position (cable TR1 to
transducer TR1 and cable TR2 to transducer TR2).

Once a key is pressed, the offset will be adjusted.

6.5 Outputs

This screen allows to program the three outputs of the instrument: Relay 1, Relay 2 and
analog output (4-20 mA).