beautypg.com

13 signal strength, 14 distance, 15 programs – Tecfluid CU Series User Manual

Page 24

background image

24

5.13 Signal

strength

This

screen allows to perform a fine adjustment of the position of the second transducer.

To do so, the second transducer must be moved to find the point where the signal level
indicated on the display is maximum.

Once this point is found

, fix the second transducer in the same way as done with the first

one in the paragraph 5.12.

5.14 Distance

If the transducers have been placed by means of the fine adjustment of signal strength
(paragraph 5.13), it is necessary to enter the practical distance at which the transducers
have been placed. This distance will be shown on the installation screen (see paragraph
5.11) as Dp.
If the fine adjustment has not been done it is not necessary to go into this screen. Simply
fix the second transducer as indicated in paragraph 5.12.
Once this step is done, exit the installation menu by pressing (Return) to go to operation
screen.

5.15 Programs

It allows to save a device programming, and retrieve it.

TRANSDUCERS INSTALLATION

If at any time you want all the transducer installation data to be stored in memory as
backup, it can be done by selecting "Save user prog. ".
To restore these data, just select "Load user prog. ".

5