Tecfluid 60M1 Series User Manual
Instructions manual, Variable area flowmeter, 60 m-1

60 M-1
Variable area
flowmeter
Series
Instructions Manual
C-MI-60M1 Rev.: 0 English Version
Technical Data
•
Accuracy:
± 2,5% of full scale value. Based on
VDE/VDI 3513
•
Scales:
In l/h, l/min, m³/h, kg/h, %, mm., etc
•
Scale length: 100 mm ± 5
•
Mounting:
Vertical (rising flow direction)
•
Total length: 220 mm ± 2
•
Pipe fittings: 1/4” BSP or NPT female thread
•
Working pressure:
10 ... 15 bar
•
Fluid temperature:
-10 ... +100ºC
The glass can support a thermal shock of 150 ºC if there is 
no internal pressure. 
The temperature difference between the interior and exterior 
of the glass tube must not exceed 80 ºC 
 
Conforms with the Pressure Equipment Directive 97/23/EC. 
 
This equipment is considered as being a pressure 
accessory and NOT a safety accessory as defined in the 
97/23/EC directive, Article 1, paragraph 2.1.3. 
 
 
The following instruction manuals are attached: 
 
60-AMD Limit Switch Instructions Manual
60-AMO Limit Switch Instructions Manual
60-AMH Limit Switch Instructions Manual
!
instrumentation
for fluids
