beautypg.com

Seaward 100 PrimeTest User Manual

Page 19

background image

PrimeTest 100

Operating Instructions

4 Specifi cation

Earth Continuity

Accuracy*

± (5% + 2 digits)

Test current

200mA minimum

Test voltage

9V nominal

Insulation resistance

Accuracy

± (5% + 2 digits)

Test

voltage

500V

Test current

>1mA into 500kΩ

Test current

<2mA into 2kΩ

Leakage Current

Accuracy

± (5% + 2 digits)

Test voltage 40V rms, 50Hz AC

Test current <5mA into 2kΩ

Cord Test

Earth continuity, insulation resistance as above.

Check for Live and Neutral open circuit, short

circuit or reversed polarity.

*When used with Seaward test lead, Part Number 347A002

Factory Set Pass/Fail limits

Class I

0.2 ohms

1.0Mohm

0.75mA

Class II

N/A

2.0Mohm

0.25mA

Cord

0.2 ohm

2.0Mohm

N/A

Earth
Continuity

Insulation
Resistance

Leakage

17