PrimeTest 100

Operating Instructions

11

If the measured value is greater than 100 ohms

no reading is shown, a cross is placed next to

the

R

PE

enunciator and a fail result is indicated

as shown below.

The test sequence is halted.

If the measured value is less than 20 ohms but

greater than the factory set pass/fail threshold,

the measured value is displayed and the unit

indicates a fail result as shown below

If the measured value is less than the factory

set pass/fail threshold, the measured value is

displayed and the unit indicates a pass result,

as shown below.