Seaward 100 PrimeTest User Manual
Page 17

PrimeTest 100
Operating Instructions
15
•
If the measured value is less than the factory set
limit a tick is placed next the R
PE
enunciator.
•
The unit will proceed with the Insulation test.
• If the Insulation Resistance is lower than the
factory set limit a cross is placed next to the
Riso enunciator and the test sequence is
halted.
• If the Insulation Resistance is greater than the
factory set limit a tick is placed next to the Riso
enunciator.
• The unit will proceed with the wiring test,
checking the live and neutral conductors for
short or open circuits or reversed connections.
• If the wiring is correct a tick is placed next to the
cord enunciator, the GOOd enunciator is
illuminated and a PASS is indicated for the
sequence.
Note: If the tested cord has a wiring fault, one
of the following enunciators will be
illuminated in place of GOOd
• OPE
n
indicates that either the live or neutral
conductor is broken (open circuit) or the
plug top fuse has blown
• Short indicates that the live and neutral
conductors are shorted together
• C
r
OSS indicates that the live and neutral
connections are crossed (live and neutral
conductors
reversed)