8-19 |
P a g e
:MR
1 - 100000
1 - 100000
Sets and/or queries the system's
Maximum Rate.
:OT?
0 = Okay
1 = Fault
Query only. Returns the Over
Temperature status.
:PE
0 or 1
0 = Disable
1 = Enable
Sets and/or queries the pulse mode
enable function.
:PM
0, 1, 2, 3
0 = CW
1 = Pulsed
2 = Burst
3 = Single
Sets and/or queries the system's Pulse
Mode.
:PW
200ns - 90%
of Rate
200ns - 90% of
Rate
Sets and/or queries the system's Pulse
Width.
:RC
1 - 5
1 - 5
Recall storage bin. Will recall the
settings in the specified user storage
bin.
:RR
0.1 - 100000
0.1 - 100000
Sets and/or queries the system's
Repetition Rate.
:SS?
16 bit decimal
Query Only. Returns the System's
State. See the System State section
below for return states.
:ST
0 or 1
0 = Disable
1 = Enable
Sets and/or queries the Start signal.
:SV
1 – 5
1 – 5
Save settings to specified storage bin.
:TB
0 or 1
0 = Disable
1 = Enable
Sets and/or queries the temperature
interlock bypass mode.
:VM?
0 – Compliance
Voltage
Query Only. Returns the Voltage
Measurement (Up to 3 decimal places).
:VN?
Ex: 1.2
Query Only. Returns the system
Version Number.
NOTES
(1)
The maximum Duty Cycle parameter allowed on the LDDC/LSC is a function of the Pulse
Mode and Period.
-In Burst Mode, the maximum allowable Duty Cycle percentage is:
-In Single Shot and Pulsed Mode, the maximum allowable Duty Cycle percentage is: