Please notify NORTEC Service if you find a fault in the system.
If there is a question please call +1 613 822 0335.
The Operator is responsible for the costs of regeneration.
When the synthesizer cartridge is received, a regenerated cartridge will be sent to
the Operator immediately.
6.3 CO
2
Bottle
bar
ba
r
ba
r
bar
3
1
4
5
2
IN
OUT
Patentierte
Reinraum-Befeuchtung
Prozessorgesteuerte
Demineralisierung
Automatisierte
Hygi enefunktionen
Modem-
Schnittstelle
Ausbeute 100%
Anbindung
Gebäudetechni k
Ein/Aus
Produkt
Gas
Synthese-
wasser
Synthese-
patrone
Reinwasser
Figure 20: Replacing the CO
2
bottle
-
Turn off the mains switch (1) of the SynPur.
-
Rotate primary shutoff valve on the bottle (2) in clockwise direction.
-
Close the valve (3) on the CO
2 bottle
by rotating it in clockwise direction.
-
Disconnect the hose (4) on the bottle.
-
Disconnect the bottle pressure reducer (5).
-
Connect the new CO
2
bottle in reverse order.
ATTENTION: After the system has been restarted it should be checked for leak-
tightness. Inspect the system occasionally over the next two days.