beautypg.com

4 material specification – Nortec SynPur User Manual

Page 24

background image

Created: 21.12.2006
Changed: 21.03.2007
Printed: 02.03.2009

page 24 of 37

4.4 Material specification


4.4.1 Electrical components


Electrical cables

Purpose:

From the SynPur to the HPS

Designation:

LIYY 7x 0.75 mm

2

(color or number code), AWG 18

Capacity:

Purpose:

From the SynPur to the GLT

Designation:

LIYY 4x0.75 mm

2

(color or number code), AWG 18

Capacity:


Standard plug with grounding connections (wall-mounted)

Purpose:

Conductivity Synthesizer, SynPur

Techn. data:
LS3400:

230 VAC, 16 A, L1, N, PE

LS3402:

120 VAC, 16 A, L1, N, PE


4.4.2 Additive

Designation:

Carbon dioxide, CO

2

Purity:

99.5%

Type of gas:

Technical gas


4.4.3 Pressure reducer


Designation:

Is fixed directly at the CO

2

-bottle


Techn. Daten:

- Back pressure 8 bar (116 psi)


- Operating pressure 6 bar (87 psi)


- The safety valve must be adjusted at 10-12 bar (145-174
psi).

-

The output, connection to console, must have a ¼“ threat