1 synpur – Nortec SynPur User Manual
Page 30
Created: 21.12.2006
Changed: 21.03.2007
Printed: 02.03.2009
page 30 of 37
6.1 SynPur
bar
ba
r
ba
r
bar
1
4
2
7
2
5
6
3
IN
OUT
Patentierte
Reinraum-Befeuchtung
Prozessorgesteuerte
Demineralisierung
Automati sierte
Hygienefunktionen
Modem-
Schnittstell e
Ausbeute 100%
Anbindung
Gebäudetechnik
Ein/Aus
Produkt
Gas
Synthese-
wasser
Synthese-
patrone
Reinwasser
Figure 18: Replacing the SynPur
-
Turn off the mains switch (3) and disconnect the plug (4).
-
Remove the two plugs (1) from the device by loosening the knurled ring on the
plug.
-
Shut off the water supply (2) on the mounting rail.
-
Open the shutoff valve (5). Wait until the system is completely depressurized.
The pressure is indicated on the pressure gage.
-
Close the shutoff valve on the CO
2
bottle (6).
-
Release the self-retracting hoses (7) (14/17 open-jawed wrench) and allow to
slide into the container.
-
Remove the SynPur from the support console and set the new device in place.
-
Connect the new device in reverse order.