reduce the overall common mode voltage, or add isolated signal conditioning between

the source and your board.

4.2.6 Isolated Grounds / Single-Ended Inputs

Single-ended inputs can be used to monitor isolated inputs, though the use of the dif-
ferential mode will increase you system’s noise immunity. The diagram below shows
the recommended connections is this configuration.

18

System with a Large Common Mode Voltage,

Connected to a Differential Input

+

-

L

arge

comm

on

m

ode v

oltag

e

be

tween

signa

l

sou

rce &

A/D

board

GND

Isolation

Barrier

When the voltage difference
between signal source and
A/D board ground is large
enough so the A/D board’s
common mode range is
exceeded, isolated signal
conditioning must be added.

+

-

Input

Amp

To A/D

A/D Board

I/O

Connector

LL GND

CH High

CH Low

10 K

10K is a recommended value. You may short LL GND to CH Low
instead, but this will reduce your system’s noise immunity.

Isolated Signal Source

Connected to a Single-Ended Input

I/O

Connector

+

-

Input

Amp

To A/D

LL GND

CH IN

A/D Board

+

-

Isolat

ed

signa

l

sour

ce