Chapter 5
Specifications
Typical for 25 °C unless otherwise specified.
Specifications in italic text are guaranteed by design.
Output relay specifications
Table 1. Output relay specifications
Number
16
Contact configuration
16 Form C (SPDT) NO, NC and Common available at connector
Contact rating
6 amperes (A) @ 240 volts AC (VAC) or 28 volts DC (VDC) resistive
Contact resistance
100 milliohms (mΩ) max.
Operate time
10 milliseconds (ms) max.
Release time
5 ms max.
Vibration
10 to 55 hertz (Hz) (Dual amplitude 1.5 millimeters (mm))
Shock
10 G (11 ms)
Dielectric isolation
500 V (1 minute)
Life expectancy
10 million mechanical operations, min.
Power on state
Not energized. NC in contact to Common.
State after RESET
button activated
Not energized. NC in contact to Common.
Isolated inputs
Table 2. Isolated inputs specifications
Number
16
Isolation
500 volts (V)
Resistance
1.6 K ohms (Ω) min.
Input high:
+5.0 VDC min. or –5.0 VDC min.
Input low:
+1.5 VDC max. or –1.5 VDC max.
DC
Input range:
! 30 VDC max.
Input high:
4.0 V
rms
min. (50-1000 Hz)
Voltage range
AC (with filter)
Input low:
1.5 V
rms
max. (50-1000 Hz)
Without filter
20 µs
Response
With filter
5 ms
Time constant
5 ms (200 Hz)
Filter control
Software programmable at each input
Power-up
Filter setting is stored on-board and will remain at the
last stored configuration
Filters
After RESET button activated
Filters off
5-1