beautypg.com

2 listen address group, 3 unlisten, 4 talk address group – Measurement Computing Digital488 User Manual

Page 68: 5 untalk, 6 local lockout, 7 device clear

background image

Section 4

IEEE 488 Primer

4.8

4.7.2 Listen Address Group (LAG)

There are 31 (0 to 30) listen addresses associated with this group. The 3

most significant bits of the data bus are set to 001 while the 5 least significant
bits are the address of the device being told to listen.

4.7.3 Unlisten (UNL)

This command tells all bus devices to Unlisten. The same as

Unaddressed to Listen. ($3F)

4.7.4 Talk Address Group (TAG)

There are 31 (0 to 30) talk addresses associated with this group. The 3

most significant bits of the data bus are set to 010 while the 5 least significant
bits are the address of the device being told to talk.

4.7.5 Untalk (UNT)

This command tells bus devices to Untalk. The same as Unaddressed to

Talk. ($5F)

4.7.6 Local Lockout (LLO)

Issuing the LLO command prevents manual control of the instrument's

functions. ($11)

4.7.7 Device Clear (DCL)

This command causes all bus devices to be initialized to a

pre-defined or power up state. ($14)

This manual is related to the following products: