Section 3 Command Descriptions

3.23

M8 will generate a Service Request when the Digital488 self-test fails.

Refer to the Test command (T0) description for details on self-tests.

M16 will generate a Service Request when the Digital488 has completed the

execution of a set of commands from the bus controller. This is used by the
controller to assure the completion of a set of commands before sending a
subsequent set of commands.

Example:

CLEAR 718

reset the Digital488

OUTPUT 718;"M4X"

select SRQ on Bus error

OUTPUT 718;"F7X"

send an invalid bus command.

Note: ERROR and SRQ LEDs should illuminate

CLEAR 718

reset the Digital488

OUTPUT 718;"M1XM4X"

select SRQ on Bus error and SRQ on Service input.

OUTPUT 718;"M5X"

This has the same effect as the command above
where M1X plus M4X equals M5X.