Table 1. Analog input specifications

Parameter

Condition

Specification

A/D converter type

Successive approximation

ADC resolution

16 bits

Sampling rate

Maximum aggregate

50 kS/s

Number of channels

8 single-ended or 4 differential; software-selectable

Input voltage range

±10 V

Working voltage

±10 V

Overvoltage protection

Power on

±30 V max

Power off

±20 V max

Input impedance

>1 GΩ

Input bias current

±200 pA, typ

INL

±1.8 LSB

DNL

16 bits no missing codes

CMRR

56 dB (DC to 5 kHz)

Input bandwidth

300 kHz

Trigger sources

Software, TRIG

Absolute accuracy (analog input DC voltage measurement accuracy)

Table 2. Absolute accuracy (no load)

Range

At full scale (typical at 25 °C)

At full scale (maximum over temperature)

System noise

±10 V

6 mV

26 mV

0.4 mVrms