32

CB-7013, CB-7033 User’s Manual

2.15 ~AAEV

Description: Enable/Disable Calibration
Syntax: ~AAEV[CHK](cr)
~

Delimiter character

AA

Address of setting module (00 to FF)

E

Command for enable/disable calibration

V

1=Enable/0=Disable calibration

Response:

Valid Command:

!AA[CHK](cr)

Invalid Command:

?AA[CHK](cr)

Syntax error or communication error may get no
response.

!

Delimiter for valid command.

?

Delimiter for invalid command.

AA

Address of response module (00 to FF).

Example:
Command: $010

Receive: ?01

Perform addreess 01 span calibration, return not enable

calibration.

Command: ~01E1

Receive: !01

Set address 01 to enable calibration, returns successful.

Command: $010

Receive: !01

Preform address 01 span calibration, returns successful.

Related Command:

Sec. 2.5,

$AA0; Sec. 2.6, $AA1

Related Topic:

Sec. 1.8,

Calibration