Firing parameters for ips e.max zirpress – Ivoclar Vivadent IPS e.max ZirPress User Manual
Page 74
74
IPS e.max Ceram on IPS e.max ZirPress
B
S
t
➚
T
H
V
1
V
2
L
Staining Technique
°C/°F
min.
°C/°F/min
°C/°F
min.
°C/°F
°C/°F
°C
ZirLiner firing – before wax-up and pressing
403/757
4:00
40/72
960/1760
1:00
450/842
959/1758
0
Stain and Characterization Firing
403/757
6:00
60/108
770/1418 1:00-2:00
450/842
769/1416
450/842
Glaze Firing
403/757
6:00
60/108
770/1418 1:00-2:00
450/842
769/1416
450/842
Add-On after Glaze Firing
403/757
6:00
50/90
700/1292
1:00
450/842
699/1290
450/842
IPS e.max Ceram on IPS e.max ZirPress/ZirCAD (Staining Technique)
IPS e.max Ceram on IPS e.max ZirPress
B
S
t
➚
T
H
V
1
V
2
L
Cut-Back and Layering Technique
°C/°F
min.
°C/°F/min
°C/°F
min.
°C/°F
°C/°F
°C/°F
ZirLiner firing – before wax-up and pressing
403/757
4:00
40/72
960/1760
1:00
450/842
959/1758
0
Wash firing (foundation)
403/757
4:00
40/72
750/1382
1:00
450/842
749/1380
0
Wash firing (foundation) characterization
403/757
4:00
40/72
750/1382
1:00
450/842
749/1380
0
1
st
Dentin / Incisal firing
403/757
4:00
40/72
750/1382
1:00
450/842
749/1380
0
2
nd
Dentin / Incisal firing
403/757
4:00
40/72
750/1382
1:00
450/842
749/1380
0
Stain Firing
403/757
6:00
60/108
725/1337
1:00
450/842
724/1335
450/842
Glaze Firing
403/757
6:00
60/108
725/1337
1:00
450/842
724/1335
450/842
Add-On with Glaze Firing
403/757
6:00
60/108
725/1337
1:00
450/842
724/1335
450/842
Add-On after Glaze Firing
403/757
6:00
50/90
700/1292
1:00
450/842
699/1290
450/842
IPS e.max Ceram on IPS e.max ZirPress/ZirCAD (Cut-Back, Layering Technique)
Firing Parameters for IPS e.max ZirPress
– Use a honey-combed tray and the corresponding pins for firing.
– Ceramic pins must not be used, since they may fuse to the restoration.
– The processing temperatures must be observed. An increase in the firing temperature results in severe vitrification
between the framework and the veneering ceramic, which may lead to cracks later. An descrease in the firing
temperature causes the ceramic to be underfired and very brittle, which may lead to delamination.
– The parameters stipulated in the Instructions for Use are coordinated with Ivoclar Vivadent furnaces (tolerance range
+/- 10°C/18°F).
– If a non-Ivoclar Vivadent furnace is used, temperature corrections may be necessary.
– Remove IPS e.max ZirPress objects from the furnace after completion of the firing cycle (wait for the acoustic signal
of the furnace).
– Allow the objects to cool to room temperature in a place protected from draft.
– Do not touch the hot objects with metal tongs.
– Do not blast or quench the objects.