Preparation of the IPS e.max ZirCAD framework (after sintering)

Adjustments of sintered IPS ZirCAD frameworks should be kept to a minimum (fine adjustment of the edges).

− For the selection of the grinding instruments, please refer to the Ivoclar Vivadent Flow Chart ”Recommended grinding

tools for IPS e.max zirconium oxide”.

− Finishing is carried out at low pressure, since too much pressure may result marginal chipping and local phase transition.

− Place IPS e.max ZirCAD framework on the model, check fit and make slight adjustments, if necessary.

− Do not ”post-separate” the bridge framework with separating disks after sintering. This may result in undesired pre-

determined breaking points, which will subsequently compromise the stability of the all-ceramic restoration.

− Check marginal areas and slightly finish, if necessary.

− For the fabrication of a pressed shoulder, the edges may be reduced up to the inner edge of the chamfer or shoulder

preparation.

− Make sure that the minimum thicknesses are maintained even after the minor adjustments.

− Before veneering, clean the framework with running water or the steam jet and dry.

− Do not blast the framework with Al

2

O

3

or glass polishing beads, since this may damage the surface.

Use the grinding instruments specifically developed for ZrO

2

to carry out the required

adjustments.

Completely sintered and finished IPS e.max ZirCAD framework.