Compensate for the shrinkage with Dentin, Transpa and Incisal materials. If required, the
gingival portions may be individualized using IPS e.max Ceram Gingiva.
Then the restoration is fired using the firing parameters for the 2
nd
Dentin and Incisal firing.
Prepare the restoration for the Stain and Glaze firing.
Apply IPS e.max Ceram Glaze, Essence and Shades.
2
nd
Dentin and Incisal firing (corrective firing)
Compensate for the shrinkage and complete the missing areas. If required, the gingival portions may be individualized
using IPS e.max Ceram Gingiva.
Firing parameters for the 2
nd
Dentin and Incisal firing
Stain and Glaze firing
Stain firing is conducted with Essence and Shades, while glaze powder or paste is used for Glaze firing. Depending on the
situation, the Stain and Glaze firings may be conducted together or separately one after the other. The firing parameters
are identical.
If necessary, the gingival portions may be characterized with IPS e.max Ceram Essence (rose, berry or aubergine)
Firing parameters for the Stain and Glaze firing – note the temperature control
IPS e.max Ceram on IPS e.max ZirPress
B
S
t
➚
T
H
V
1
V
2
L
Gingiva Technique
°C/°F
min.
°C/°F/min
°C/°F
min.
°C/°F
°C/°F
°C/°F
Stain firing
403/757
6:00
60/108
725/1337
1:00
450/842
724/1335
450/842
Glaze firing
403/757
6:00
60/108
725/1337
1:00
450/842
724/1335
450/842
IPS e.max Ceram on IPS e.max ZirPress
B
S
t
➚
T
H
V
1
V
2
L
Gingiva Technique
°C/°F
min.
°C/°F/min
°C/°F
min.
°C/°F
°C/°F
°C/°F
2
nd
Dentin and Incisal firing
403/757
4:00
40/72
750/1382
1:00
450/842
749/1380
450/842