The following points should be observed for ceramic furnaces, used for the crystallization of IPS e.max CAD:
– Crystallization should be carried out in an Ivoclar Vivadent ceramic furnace (e.g. Programat P300, P500, or P700).
– If you would like to use other, untested ceramic furnaces, please consult Ivoclar Vivadent about their compatibility with
IPS e.max CAD.
– Basically, the following applies:
Ceramic furnaces without
– function for controlled (long-term) cooling
– vacuum function
cannot be used.
– Before the first crystallization and every six months after that, the ceramic furnace must be calibrated.
– Depending on the mode of operation, more frequent calibration should be required. Follow the instructions for the
manufacturer.
The following aspects should be observed for conducting the crystallization:
– Use only IPS Object Fix Putty or Flow as an auxiliary firing paste to place the restoration directly on the IPS e.max CAD
Crystallization Tray.
– IPS e.max CAD restorations must not be directly placed on the IPS e.max CAD Crystallization Tray and Pins, e.g. without
auxiliary firing paste, for crystallization.
– Use only the IPS e.max CAD Crystallization Tray and the corresponding Pins, since they store the heat necessary for slow
and above all tension-free cooling of the glass-ceramic.
– Always conduct the crystallization under vacuum.
– Remove IPS e.max CAD objects from the furnace after completion of the firing cycle (wait for the acoustic signal of the
furnace).
– Allow the objects to cool to room temperature in a place protected from draft.
– Do not touch the hot objects with metal tongs.
– Do not blast or quench the objects.
62
e.max
®
CAD
–
Crystallization Parameters
IP
S
Furnace
P80
P100
P200
P300
P500
P700
PX1
EP 600
EP 3000
EP 5000
Closing time
S
min.
6:00
6:00
6:00
6:00
6:00
6:00
Stand-by
temperature
B
°C/°F
403/757
403/757
403/757
403/757
403/757
403/757
Heating rate
t
1
°C/°F/min
–
60/108
60/108
60/108
60/108
60/108
Firing
temperature
T
1
°C/°F
–
770/1418
770/1418
770/1418
770/1418
770/1418
Holding time
H
1
min.
–
0:10
0:10
0:10
0:10
0:10
Heating rate
t
2
°C/°F/min
30/54
30/54
30/54
30/54
30/54
30/54
Firing
temperature
T
2
°C/°F
850/1562
850/1562
850/1562
850/1562
850/1562
850/1562
Holding time
H
2
min.
10:00
10:00
10:00
10:00
10:00
10:00
Long-term
cooling
L
°C/°F
700/1292
700/1292
700/1292
775/1427
1:30 min
700/1292
0:20 min
700/1292
700/1292
Cooling rate
t
l
°C/°F/min
–
–
0
–
–
0
Vacuum 1
1
1
1
2
°C/°F
–
550/1022
770/1418
550/1022
770/1418
550/1022
770/1418
550/1022
770/1418
550/1022
770/1418
Vacuum 2
2
1
2
2
°C/°F
550/1022
850/1562
770/1418
850/1562
770/1418
850/1562
770/1418
850/1562
770/1418
850/1562
770/1418
850/1562
Crystallization of IPS e.max CAD MO
Only the program memories 65-69 can be used with the Programat P100.
Please note:
– IPS e.max CAD HT and LT may also be crystallized with this program.
– If only max. two restorations are crystallized at the same time, the closing time S can be reduced to 1:30 min.