Crystallization – Ivoclar Vivadent IPS e.max CAD Labside User Manual
Page 36

36
Veneer and anterior crown on dies
Anterior crown on ZrO
2
abutment (Straumann
®
Anatomic IPS e.max
®
Abutment)
Do not place the IPS e.max CAD restoration on metal pins and do not use a honey-combed
firing tray
Slightly overfill the restoration cavity with IPS Object Fix Putty or Flow
Place the restoration on the IPS e.max CAD Crystallization Tray
Place the firing tray in the furnace and start the crystallization with the respective parameters
Crystallization
The following points should be observed for the crystallization:
– Crystallization should be carried out in an Ivoclar Vivadent furnace (e.g. Programat P300, P500 or
P700).
– Use only IPS Object Fix Putty or Flow as an auxiliary firing paste to ideally place the restoration on the
firing tray.
– Slightly overfill the restoration with the auxiliary firing paste so that there is a reservoir.
– Use only the IPS e.max CAD Crystallization Tray, since it stores the heat necessary for slow and above all tension-free
cooling of the restoration.