CMG-5T Compact
single-ended output for 1 g (≈ 9.81 m/s
2
) input acceleration. The
standard frequency pass band is flat to acceleration from DC to 100 Hz
(although other low pass corners from 50 Hz to 100 Hz can be ordered.)
A high frequency option provides flat response from DC to 200 Hz.
Each accelerometer is delivered
with a detailed calibration sheet
showing its serial number,
measured frequency response,
sensor DC calibration levels and
the transfer function in
poles/zeroes
notation.
Installation is simple, using a
single fixing bolt to attach the
sensor to a hard surface. If
required, you can also level the
sensor using its two adjustable
levelling feet.
Optionally, you can use a Güralp Hand-held Control Unit (HCU) and
breakout box to distribute power and calibration signals to the sensor
and to receive the signals it produces. It is available in standard, rack-
mounted and water-resistant portable formats.
The accelerometer housing
itself is completely
waterproof, with a hard
anodised aluminium body
and “O”-ring seals
throughout.
CMG-5T
Compact instruments have
been tested for long periods
of total immersion in water.
4
Issue E