Operator's Guide

The current offset values will display. Once the process has completed

the Scream! status window will display (in green):

CENTERING COMPLETE

Note: if you enter the command CENTRE or CENTER the process will

still take place but the values will not display.

2.4.3

Removing offsets using a third party digitisers

To remove offsets while using third party digitisers, connect pin U on
the sensor connector to pin Y (see section 5 on page 24). Note that the

sensor must be powered while performing this operation.

The connection between pins U and Y must be maintained for seven
seconds before the centring function is activated. The connection

must then be released. Centring will continue for as long as is
required after the connection is broken,

2.5

The hand-held control unit

The 5T Compact can be supplied with an

optional hand-held control unit (HCU).
This is a portable device which provides

easy access to the seismometer's controls,
as well as displaying the output

acceleration on an analogue meter. The
5T Compact uses the same HCU as Güralp

5TB and 40T instruments but there are a
few differences. The output from the 5T

Compact is acceleration not velocity so, to
view the high gain acceleration, select the

'HIGH VEL' switch and to select the low
gain acceleration select the 'LOW VEL'.

The 5T Compact also doesn't output the
mass position and has no electronic

centring so these parts of the HCU should
be ignored.

2.5.1

Signal meter

The upper section of the HCU contains a simple voltmeter for

monitoring the signals from the instrument.

To monitor the low-gain outputs, switch the dial to V, N/S or E/W
LOW VEL
according to the component you want to monitor.

May 2011

11