beautypg.com

2 electrical connections – Guralp Systems CMG-5T Compact User Manual

Page 22

background image

CMG-5T Compact

The normalising constant A is calculated at a normalising

frequency value fm = 1 Hz, with s = j fm, where j = √–1.

H(s) is the transfer function of the sensor, which can be expressed
in factored form:

In this equation z

n

are the roots of the numerator polynomial,

giving the zeroes of the transfer function and p

m

are the roots of

the denominator polynomial, giving the poles of the transfer
function.

In the calibration pack, G is the sensitivity given for each component on

the first page, whilst the roots z

n

and p

m

, together with the normalising

factor A, are given in the Poles and zeroes table. The poles and zeroes

given are measured directly at Güralp Systems' factory using a spectrum
analyser. Transfer functions for the vertical and horizontal sensors may

be provided separately.

4.2

Electrical connections

Each channel inside the 5T Compact sensor has four output lines: a pair
of differential outputs with low gain and another pair with high

(nominally 10 ×) gain.

Optionally, the second gain block can be configured at the factory to act
as a high-pass filter to remove the DC offset at the output terminal. This

filter can have a corner frequency of 0.05 Hz (20 s) or 0.025 Hz (40 s). In
these instruments, the gain at the additional outputs is set to × 1 (unity).

The output offsets of a high-pass filtered output cannot be zeroed using
the DC offset adjustment screws; in any case, this offset should not be

more than ± 1 mV. The high-pass filtered output is likely to take around
5 times the time constant of the high-pass circuit to settle down. This

time constant is detailed on the calibration sheet, together with
accurately-measured frequency values.

The two pairs of output lines are balanced about signal ground so that

either differential drive or single-ended drives of opposite polarity
(phase) are available. For a single-ended drive, the signal ground must

be used as the signal return path. You must not ground any of the active
output lines, as this would allow damaging currents to flow through the

output circuits. Also, if single-ended outputs are used, the positive
acceleration outputs must be the ones interfaced to the recorder.

22

Issue E