beautypg.com

3 calibrating accelerometers – Guralp Systems CMG-5T Compact User Manual

Page 14

background image

CMG-5T Compact

signal into the sensor. This can be either a sine wave, a step function or

broad-band noise, depending on your requirements. As well as going to
the sensor, the calibration signal is returned to the digitizer on a full rate

channel (older digitizers used one of the 4 Hz auxiliary (Mux) channels).
The calibration signals and sensor output all travel down the same cable

from the sensor to an analogue input port on the digitizer.

The signal injected into the sensor gives rise to an equivalent
acceleration
(EA on the above diagram) which is added to the measured

acceleration to provide the sensor output. Because the injection circuitry
can be a source of noise, a Calibration enable line from the digitizer is

provided which disconnects the calibration circuit when it is not
required. Depending on the factory settings, the Calibration enable line

must be either allowed to float high (+5 to +10 V) or held low (0V,
signal ground) during calibration: this is specified on the sensor's

calibration sheet.

The equivalent acceleration corresponding to 1 V of signal at the
calibration input is measured at the factory and can be found on the

sensor calibration sheet. The calibration sheet for the digitizer
documents the number of counts corresponding to 1 V of signal at each

input port.

The sensor transmits the signal differentially, over two separate lines.
This improves the signal-to-noise ratio by increasing common mode

rejection. If you are not using a Güralp Systems digitizer, the voltage
across the output pins should be halved before converting to

acceleration.

CMG-5T Compact instruments are tuned at the factory to produce 1 V of
output for 1 V input on the calibration channel. For example, a sensor

with an acceleration response of 0.25 V/ms

-2

should produce 1 V output

given a 1 V calibration signal, corresponding to 1/0.25 = 4 ms

-2

= 0.408 g

of equivalent acceleration.

3.3

Calibrating accelerometers

Both the DM24 digitizer and Scream! software allow direct configuration
and control of any attached Güralp instruments. For full information on

how to use a DM24 series digitizer, please see its own documentation. If
you are using a third-party digitizer, you can still calibrate the

instrument as long as you activate the Calibration enable line correctly
and supply the correct voltages.

1. In Scream!'s main window, right-click on the digitizer's icon and

select Control.... Select the Calibration tab.

2. Under Component, select ALL. The CMG-5T has a single

14

Issue E