Exposure and rate resolution and limits, Appendix c, C.1 exposure and rate resolution and limits – Fluke Biomedical 8000 Victoreen User Manual
Page 89

Appendix
Exposure and Rate Resolution and Limits
C
C-1
Appendix C
Exposure and Rate Resolution and Limits
C.1 Exposure and Rate Resolution and Limits
Exposure and Rate Resolution, Standard Units (R)
Ion
Low Sensitivity
High Sensitivity
Chamber
Exposure Rate Exposure Rate
Mammo
.01 R
1 R/min
1 mR
.1 R/min
Fluoro
.1 mR
.01 R/min
.01 mR
1 mR/min
R/F
1 mR
.1 R/min
.1 mR
.01 R/min
Scatter
.1 mR
.01 R/min
.01 mR
1 mR/min
Internal
1 mR
.1 R/min
.1 mR
.01 R/min
Rate Limits, Standard Units (R)
Ion Chamber
Low
Sensitivity
High
Sensitivity
Mammo
10 kR/min
1000 R/min
Fluoro
250 R/min
25 R/min
R/F
1000 R/min
100 R/min
Scatter
100 R/min
10 R/min
Internal
1000 R/min
100 R/min
Exposure and Rate Resolution, SI Units (Gy)
Ion
Low Sensitivity
High Sensitivity
Chamber Exposure
Rate
Exposure
Rate
Mammo
.1 mGy
.01 Gy/min
.01 mGy
1 mGy/min
Fluoro
1 uGy
.1 mGy/min
.1 uGy
.01 mGy/min
R/F
.01 mGy
1 mGy/min
1 uGy
.1 mGy/min
Scatter
1 uGy
.1 mGy/min
.1 uGy
.01 mGy/min
Internal
.01 mGy
1 mGy/min
1 uGy
.1 mGy/min