Fluke Biomedical 8000 Victoreen User Manual
Page 47

Operation
Modes of Operation
2
2-35
0
mR
Please wait…
HVL MAMMO
INTEG
HIGH
MODE CHAMBER
MODE SENS
Please wait while the NERO mAx prepares to take an exposure. If the NERO mAx detects a fault
condition such as a high offset, an error message is displayed and the user must correct the error to
continue (see Section 4.3--Error Messages--for more information). When the NERO mAx is ready to take
an exposure it will beep and prompt for an exposure.
0
mR
MAKE EXPOSURE # 01
HVL MAMMO
INTEG
HIGH
MODE CHAMBER
MODE
SENS
Make the first exposure with no filtration in the beam.
368
mR
Please wait…
HVL MAMMO
INTEG HIGH
MODE CHAMBER
MODE SENS
REJECT
The exposure data from the first exposure is displayed and the NERO mAx prepares to take the second
exposure. Please wait while the NERO mAx prepares to take the next exposure. If the NERO mAx
detects a fault condition such as a high offset, an error message is displayed and the user must correct
the error to continue (see Section 4.3--Error Messages--for more information).
368
mR
MAKE EXPOSURE # 02
HVL MAMMO
INTEG
HIGH
MODE CHAMBER
MODE
SENS
REJECT
The NERO mAx is now ready to take the next exposure. Position an aluminum absorber in the beam and
make the exposure. If the previously measured exposure is not acceptable, press the reject softkey to
retake the exposure. If an overrange occurs, an error message is displayed and the exposure must be
taken over.