Fluke Biomedical 8000 Victoreen User Manual
Page 33

Operation
Modes of Operation
2
2-21
MAKE EXPOSURE
FILTER = 50 - 100
FLUORO CONT
HIGH
MODE TYPE SENS
Start the fluoroscopic exposure.
4.49
R/min
4.1
mA
FLUORO CONT
HIGH
MODE TYPE SENS
SAVE
After the exposure starts, the NERO mAx displays exposure rate and mA. When the desired exposure
rate and/or mA are displayed, press the SAVE key to acquire kV data.
4.49
R/min
4.1
mA
FLUORO CONT
HIGH
TURN OFF
MODE TYPE SENS
BEAM
After kV data is acquired, the NERO mAx prompts to turn off the beam.
4.49
R/min
4.1
mA
FLUORO CONT
HIGH
ANALYZE
MODE TYPE SENS
DATA
After the beam is turned off, the NERO mAx analyzes the fluoro exposure kV data.
89.6
kVp Avg
87.8
kV Eff
4.34
R/min
91.9
kV Peak
4.1
mA
FLUORO CONT
HIGH
Please
MODE TYPE SENS
Wait…
After data analysis is complete, kV, exposure rate and mA are displayed. If an overrange is detected an
error message is displayed. If the calculated kV is above or below the selected filter kV range, “High” or