Fluke Biomedical 8000 Victoreen User Manual
Page 77

4
Troubleshooting
Waveforms-Partial kV Waveforms
4-15
In the
exposure
below, part of the kV waveform appears to disappear after an initial peak. This is because after the initial
peak, the radiation output drops below the level at which the NERO mAx can accurately reproduce the kV
waveform. This level is approximately 400 counts on the channel A radiation waveform.
kV Waveform
High Sensitivity
0
5
10
15
20
25
30
35
40
45
0
2
4
6
8
10
12
14
mSec
kV
kV
kV Waveform
0
10
20
30
40
50
60
70
80
90
0
20
40
60
80
100
120
140
160
180
mSec
kV
kV
Radiation Waveform
0
100
200
300
400
500
600
700
800
900
0
20
40
60
80
100
120
140
160
mSec
counts
Channel A
Channel B
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)