Fluke Biomedical 8000 Victoreen User Manual
Page 53

Operation
Modes of Operation
2
2-41
0.0
R
Please Wait…
HVL CT
HIGH 10
mm
MODE CHAMBER
SENS
BEAM
Please wait while the NERO mAx prepares to take an exposure. If the NERO mAx detects a fault
condition such as a high offset, an error message is displayed and the user must correct the error to
continue (see Section 4.3--Error Messages--for more information). When the NERO mAx is ready to take
an exposure it will beep and prompt for an exposure.
0.0
R
MAKE EXPOSURE # 01
HVL CT
HIGH 10
mm
MODE CHAMBER
SENS
BEAM
Make the x-ray exposure with no filtration in the beam.
6.71
R
Please Wait…
HVL CT
HIGH 10
mm
MODE CHAMBER
SENS
BEAM
REJECT
While the exposure from the first exposure is displayed, the NERO mAx prepares to take the second
exposure. Please wait while the NERO mAx prepares to take the next exposure. If the NERO mAx
detects a fault condition such as a high offset, an error message is displayed and the user must correct
the error to continue (see Section 4.3--Error Messages--for more information).
3.43
R
MAKE EXPOSURE # 02
HVL CT
HIGH 10
mm
MODE CHAMBER
SENS
BEAM
REJECT