Fluke Biomedical 8000 Victoreen User Manual
Page 39

Operation
Modes of Operation
2
2-27
250.9
mR/frame
45.6
mAs/frame
AMSE LOW
MODE SENS
SAVE
After the exposure starts, the NERO mAx displays exposure/frame and mAs/frame. When the desired
exposure rate is displayed, press the SAVE key to acquire kV data.
250.9
mR/frame
45.6
mAs/frame
AMSE LOW
TURN OFF
MODE SENS
BEAM
After kV data is acquired, the NERO mAx prompts to turn off the beam.
250.9
mR/frame
45.6
mAs/frame
AMSE LOW
ANALYZE
MODE SENS
DATA
After the beam is turned off, the NERO mAx analyzes the fluoro exposure kV data.
65.6
kVp Avg
57
msec/frame
65.0
kV Eff
250.9
mR/frame
67.2
kV Peak
45.6
mAs/frame
AMSE LOW
Please
MODE SENS
Wait…
After data analysis is complete, kV, exposure rate, mAs/frame and time/frame are displayed. If an
overrange is detected an error message is displayed. If the calculated kV is above or below the selected
filter kV range, “High” or “Low” is displayed instead of kV (see Section 4.3--Error Messages--for more
information). After the exposure data is displayed, the NERO mAx displays a "Please wait..." message as
it prepares for the next exposure.