3 specifications – Fluke Biomedical 8000 Victoreen User Manual
Page 8

Victoreen 8000
Operators Manual
1-4
1.3 Specifications
Kilovoltage
Measured during the first 480 ms of exposure
Accuracy:
0.5 kV or ± 1%
Reproducibility:
0.5 kV or ± 1%
Range:
W/Al
30 - 60 kV
50 - 100 kV
80 - 160 kV
Mo/Mo
22 - 35 kV
Mo/Rh
22 - 40 kV
Mo/Al
22 - 49 kV
Rh/Rh
25 - 49 kV
Rh/Al
25 - 49 kV
Time To Display
Radio & Mammo: 3 seconds for 0.1 second exposure
1 second for each 32 ms of exposure time
Fluoro & AMSE: 15 seconds for all exposures
Time
Measured during entire exposure at 90% rise/fall of waveform.
Accuracy: 1
ms
Range:
1 ms to 60 sec
Exposure/Exposure & Rate
Measured during entire exposure; kVp corrected.
Accuracy: ±
5%
Reproducibility:
(Radio & Mammo Modes) ± 2% or 2 mR
Range:
1 mR minimum
mAs and mA
Measured during entire exposure
Accuracy: ±
2%
Reproducibility:
± 1% or 0.2 mAs
Range:
1 - 1000 mA
HVL
Accuracy: ±
5%
Range:
.1 - 99.9 mmAl
Physical
Display:
240 x 60 pixel, super twist LCD w/ccfl backlight
Power:
115 or 230 VAC External Supply. Rechargeable internal batteries supply more than
4 hours of continuous service with overnight charge.